SPI
FM EMC test results concerning the PSD subassembly
Edition
1, Revision 0 (12/3/2001)
Author:
J. Knödlseder
1.
INTRODUCTION
This
document provides the results of the SPI FM EMC tests that concern the PSD
subassembly. The EMC tests have been performed by CNES during the period
13/2/2001 - 21/2/2001 at INTESPACE. During these tests, a number of anomalies
have been discovered tha
t
concern the PSD subassembly :
- NCR
963 : PSD susceptibility when CDE is switched-on
- NCR
964 : PSD susceptibility at 30 MHz RS 4V/m
The
aim of this document is to summarise the results of a deeper investigation of
these anomalies
which is based on the TM data that has been accumulated during the EMC tests.
In addition, the impact on the scientific PSD performance is also studied.
The
following quantities that were extracted from the TM are of importance for PSD
performance verification in different electromagnetic environments :
- Number
of PSD time-tags : allows the detection of additional noise triggers which
could arise from noise on some of the PSD channels
- PSD
channel rates : allows to identify noisy channels
- Number
of pure PSD
events (PP) : allows to identify PSD noise triggers that are not associated
with a AFEE time-tag
- PSD
baseline noise : allows an estimation of the PSD performance drop (see RD1 for
the definition of the baseline noise)
- PSD
pulse shape visualisation : helps to identify the reason for enhanced PSD
triggers
While
the first two points were monitored and verified by CNES during the EMC tests,
all points were investigated by CESR using a post-processing of the EMC TM data.
2.
REFERENCE
DOCUMENTS
RD1 PSD
Software
description, SPI-NT-4232-4194-CESR, Ed. 2, Rev. 0, 06/04/00
RD2 PSD
User Manual, SPI-MU-423-4215-CESR, Ed. 2, Rev. 0, 19/02/01
RD3 EMC
FM test procedure, SPI-PR-0-16505-CNES, Issue 1, Rev. 1, 12/02/01
3.
TEST
RESULTS
Fig.
1: Typical noise trigger on channel 11. The oscillation is due to the front-end
reset of the PSD which eventually may lead to a self-triggering of the PSD
sub-assembly.
3.1
Channel 11 noise triggers and FET setting
During
previous SPI performance verifi
cation
tests, it turned out that the trigger threshold on PSD channel 11 is relatively
low compared to the other thresholds (recall that only one parameter allows the
setting of the FET trigger threshold, although all 19 PSD channels have a
separate imple
mentation
of the front-end trigger logic; see RD2). However, since a too high FET
threshold might compromise the scientific performance of PSD, it has been
decided to configure PSD during EMC tests with a relatively low threshold (FET
= 5), although some
noise
triggers may be expected in this case on channel 11.
Figure
1, which shows a PSD pulse shape for channel 11, illustrate such a typical
noise trigger. Indeed, the pulse shape that is seen in case of a noise trigger
always resembles the shape shown in Fig. 1, and it turned out that this noise
is produced by P
SD
itself. In fact, after a PSD front-end trigger, the entire front-end is reset,
leading to a
n
electrical perturbation on the PSD input lines. This perturbation is a decaying
oscillation which
reflects the front-end reset signal. Eventually, if the front-end trigger
threshold is too low, PSD may trigger on this reset signal. Since channel 11
has effectively the lowest front-end threshold, it is the first channel that
becomes sensitive to self-
triggering.
This behaviour needs to be kept in mind to understand the following test results.
Typically,
during the EMC tests, about 50% of the channel 11 triggers were noise triggers
(i.e. the channel 11 trigger rate was enhanced by a factor 2 with respect
to the other 18 PSD channels).
3.2.
PSD susceptibility when CDE is switched-on (NCR 963)
CNES
reported that the channel rate of channel 11 increased after the switch-on of
the CDE. Following RD3, the CDE was switched-on the 14/2/2001 at 10:30. However,
the
investigation of the EMC TM did not reveal any change in the channel 11 rate at
this moment.
This is illustrated in Fig. 2 which shows the channel 11 rate (in units of
counts / 64 seconds) as function of time (indicated by RowNumber). The drop of
channel rate to 0 corre
sponds
to the period where SPI was switched into CONF mode and the HT were switched
off. Apparently, the channel 11 rate with CDE on is similar to the rate with
CDE off.
Fig.
2: Variation of PSD channel 11 rate as function of time during the transition
CDE off / CDE on (in units of counts / 64 seconds).
3.3.
PSD susceptibility from 80 MHz - 120 MHz (NCR 963)
CNES
reported that the channel rates 5, 11, 14, 15, and 17 decreased especially at
80 MHz (NCR 963). This behaviour has been observed during the nominal RS +y/-z
tests, conducted the 17/2/2001 from 11:04 - 11:37.
Figure
3. shows the channel rates for all 19 PSD channels (top) and the LLD rate
(bottom) for this test period (the injection frequencies are indicated in the
lower panel). Three phenomena can be observed:
- At
30 MHz, the LLD rate rises from 1500 to 6500 triggers per second. At the same
time, the channel 11 rate rises from 12 to 16 counts/sec. Also for channel 4, a
slight increase is seen from 6 to 7 counts/sec.
- At
80 MHz, the channe
l
rates for channels 5, 9,
14,
15, and 17 are considerably reduced. The LLD rate is also reduced. Attenuating
the 80 MHz signal by -6 db brings the rates back to their nominal values.
- At
80 - 100 MHz, the channel 11 rate and the LLD rate are reduced. Attenuating the
80 MHz signal by -6 db
brings the rate back to almost the nominal values.
From
the analysis of the TM recorded on the CESR workstation the following
explanations were obtained :
1. At
30 MHz, the noise in the PSD pulse shapes is slightly enhanced on channel 4 and
11, showing a 30 MHz oscillation in the recorded pulse shapes (see also section
3.4). The additional noise adds to the front-end reset noise that is present on
all channels (see section 3.1), which then brings also
channel 4 in the regime where it sees the PSD front-end perturbations (and
channel 11 sees even more such perturbations).
However,
it is not yet clear why channel 4 shows this susceptibility since it has not a
particularily low threshold level.
Probably,
the source of the 30 MHz susceptibility is not located in the PSD, and maybe
the PA2 of channel 4 is particularily susceptible to 30 MHz perturbations.
Note:
During the redundant RS +y/-z tests (on 19/2/2001 from 17:21 - 17:48) only a
slight increase in the LLD rate from 1600 to 2100 triggers per second has been
observed at 30 MHz. Additionally, the channel rates did not increase.
2. For
some channels, the
80
MHz injection leads to an effective increase
of
the FET trigger threshold
.
This is illustrated in Fig. 4 which shows the lower threshold of PSD events
(PE) for all 19 detection channels and different RS +y/-z injection
frequencies. Apparently, channels 5, 14, and 15 show a considerably enhanced
threshold which effectively reduces the channel rates (since a higher threshold
means less PSD triggers). Channel 9 and 17 suffer the same problem, although
this trend i
s
not clearly visible in Fig. 4 (the counting statistics for the threshold
determination was very poor, hence the threshold determination is not very
accurate).
The
prove that it is indeed the front-end trigger (FET) threshold that is
effectively increased by a 80 MHz injection comes from the analysis of the
average PSD pulse shapes (see Fig. 6). Apparently, at 80 MHz and also slightly
at 90 MHz, the pulse shape
is
seriously truncated at the start of the pulse, indicating that the pulse trigger
s
came too lat
e.
The pulse trigger, however, is determined by the FET threshold
setting, and truncated pulses are a clear sign of a too high FET level.
It
is interesting to note that except for ch
annel
9
the
concerned detectors are all in the same region of the detection plane
(direction -y/-z). Yet channel 16, which is in the same area, does not show a
similar behaviour. Additionally, during the RS -y/-z tests where the radiation
emitter was closer to the susceptible detectors, no 80 MHz susceptibility has
been detected.
Note:
During the redundant RS +y/-z tests (on 19/2/2001 from 17:21 - 17:48) a similar
reduction of the channel rates 5, 9, 14, 15, and 17 was observed, and the PE
threshold showed a comparable increase.
Fig.
3: Channel rates (top) and LLD rate (bottom) during the RS +y/-z injection from
30 MHz to
120
MHz. For the 80 MHz injection, the channel rates drop for channels 5, 9, 14,
15, and 17.
3. Channel
11 generally shows a substantial amount of noise triggers due to the relatively
low front-end threshold for this channel (see section 3.1). A slight increase
of the FET, as observed for channels 5, 14, and 15, would bring channel 11 out
of the noise and would effectively reduce the trigger rate. Indeed, at 100 MHz,
the number of noise triggers on channel 11 is reduced from about 50 % to 30%,
which can perfectly explain the reduction of the channel
11
rate from 12 counts / sec to about 8 counts / sec.
However, with the poor statistics of the accumulated data (only about 3 minutes
for each injection) it is difficult to detect a small variation in the PSD
event threshold. Thus, the hypothesis of an increased FET threshold as origin
of the channel 11 rate
drop could not be verifie
d.
Note:
During the redundant RS +y/-z tests (on 19/2/2001 from 17:21 - 17:48), channel
11 did not show any decrease in the channel rate at 80 MHz. However, reductions
of the channel rates were seen at 50 MHz, 70 MHz, and 90 MHz.
Fig.
5: PSD event lower threshold (in PHA channels). At 80 MHz, the threshold of
channels 5, 14, and 15 is considerably enhanced.
Fig.
6: Average pulse shape for channel 5 for various RS +y/-z injection
frequencies. Between
80
- 90 MHz the start of the pulse
shapes
is truncated.
3.4.
PSD susceptibility at 30 MHz during RS -y/-z injection (NCR 964)
CNES
reported a susceptibility at 30 MHz (NCR 964) which has been observed during the
RS
-y/-z injection, conducted the 16/2/2001 at 16:55.
Fig.
7: PSD trigger on 30 MHz oscillation for channel 4.
Indeed,
the analysis of PSD pulse shapes recorded during this period reveals a strong
30 MHz oscillation superimposed on the detector pulses for some of the PSD
channels. Figure 7 shows such an oscillation, recorded on c
hannel
4. This oscillation leads to an important increase of the channel rate for the
susceptible channels due to noise triggers. From the analysis of the P
SD
pulse shape noise (see Fig. 8) it has been determined that channels 4 and 11
are the most susceptible ones. Also channels 0, 6, 13, and 18 show some noise
increase at 30 MHz, while the other channels seem to be almost unaffected by
the 30 MHz injection. Reducing the injection amplitu
de
by 6 db considerably reduces the noise on the PSD pulse shapes, br
inging
the trigger rate for channel 4 and 11 into a reasonable range.
Since
most of the PSD electronic is common for all 19 detection channels (see RD2),
the susceptibility difference among the 19 PSD channels makes it questionable
if the PSD subassembly itself is susceptible to the noise injection. It could
also be that so
me
of the pre-amplifiers are particularly susceptible to a 30 MHz signal, and PSD
simply sees the noise of those pre-amplifiers.
Note:
The 30 MHz susceptibility detected during the RS +y/-z injections is much
weaker than the RS -y/-z susceptibility. On the other hand, the RS -y/-z inje
ction
on the redundant
SPI
has shown a similar susceptibility on channel 4 and 11 than on the main SPI.
3.5.
PSD susceptibility at 50 MHz during redundant RS +y/-z injection
During
the RS +y/-z injection on the redundant SPI (19/2/2001 from 17:21 - 17:48),
enhanced noise has been found at 50 MHz on PSD channels 1, 3, and 5. A
inspection of the PSD pulse shapes recorded during this period has revealed
that the 50 MHz signal propagated into the PSD input channels. The mean noise
observed on channels 1, 3, and 5 was 4.0 - 4.3 digitalisation units, roughly
30% higher than the normal noise (see also Fig. 8).
Fig.
8: PSD pulse shape noise (in PSD digitisation units) as function of the RS
-y/-z injection frequency. Some channels show enhanced noise at a frequency of
30 MHz (in particular channels 4, 11, 6, and 13.
4.
CONCLUSIONS
The
analysis of the TM recorded during the EMC test leads to the following
conclusions concerning the susceptibility of the PSD sub-assembly:
- The
CDE status (off or on) has no impact on the PSD sub-assembly. NCR 963, raised
during the EMC tests, was probably based on an erroneous comparison of channel
11 trigger rates.
No
CDE susceptibility has been found.
- The
PSD sub-assembly s
hows
some susceptibility
at
80 MHz which effectively raises the front-end threshold (FET) on some of the
PSD channels (in particular channel 5, 9, 14, 15, and 17). From an electronics
standpoint, this susceptibility is not yet understood.
- Between
80 - 100 MHz the noise trigger rate of channel 11 is reduced, probably also due
to a slight increase of the front-end threshold.
- The
system PA2 - PSD shows some susceptibility at 30 MHz due to a propagation of the
30
MHz signal into the PSD trigger electronics for some of the PSD channels (in
particular channel 4, 11; also less for channels 0, 6, 13, and 18).
This
susceptibility is much stronger for an injection on the -y/-
z
side than for the +y/-z side.
It is not yet clear i
f
this susceptibility appears in the PSD sub-assembly, or if it appears on some
pre-amplifiers or the cable that connects the pre-amplifiers to the PSD
sub-assembly
. - At
50 MHz, a small susceptibility similar to the 30 MHz susceptibility has been
observed on channels
1, 3, and 5.
However, it is weaker than the 30 MHz -y/-z susceptibility.
The
80 MHz susceptibility will potentially make if difficult to have a stable lower
energy threshold for all 19 PSD channels. Additionally, an increase of the
threshold will truncate PSD pulses at the start (see Fig. 4), leading to a
modifica
tion
of the PSD pulse characteristics. In this case, the background rejection
efficiency of the PSD will be reduced since the measured pulses shapes will
deviate significantly from the uplinked template library. It is therefore
essential to monitor the PS
D
pulse characteristics continuously throughout the mission. Also, a variation of
the PSD lower energy threshold on some channel may be used as indication of a
potential decrease of PSD efficiency on this channel.
The
30 MHz susceptibility implies additiona
l
noise on the detector pulses for some channels. Since it is more difficult to
distinguish a single-site from a multiple-site interaction with such additional
noise, it is expected that the background rejection efficiency will be degraded
for these chann
els
in case of 30 MHz noise.
For
50 MHz, the situation is similar, but the amplitude of the perturbation is
smaller than at
30
MHz.